Published 11 November 2026 at 18:19
A caliper falls off the bench, gets squeezed in a drawer, or is used for months on oiled material without being cleaned. It still looks fine, the display shows neat numbers, and the jaws close. But nobody knows whether it reads correctly until a customer files a complaint or a part doesn’t fit in the assembly. That is exactly the situation a check set is made for.
What the set contains
The set is designed for checking calipers according to DIN 862. It comprises three steel gauge blocks of 30 mm, 41.3 mm and 131.4 mm in grade 1 according to ISO 3650, two setting rings of Ø4 mm and Ø25 mm, and a pair of gloves. Everything comes in a wooden case together with the manufacturer’s measurement report. The uneven lengths are chosen so that you check the scale at several points of the measuring range rather than only at round numbers. The gloves are included because the warmth of your hands makes steel expand enough to distort a check.
What problems it solves
You get a fixed, traceable basis for deciding whether a caliper is still within tolerance. The blocks check the outside measurement, and the rings check the inside measurement with the small jaws. These are two different sources of error, and they can deviate independently of each other. Without references, your confidence in the instrument rests on habit and experience. With the set, it rests on a figure you can write down in a checklist.
When and why it is used
Typical occasions are when putting a new caliper into service, after a drop or a suspicious measurement, and at fixed intervals within the quality system. The procedure is simple: clean the jaws, place the block between them with normal measuring force, read the value and compare it with the nominal value. The deviation is noted. If it lies outside what you accept, the instrument is taken out of use or sent for adjustment.
Three concrete situations
The machine shop: Several turners share a set of calipers. A weekly check against the blocks takes a few minutes and reveals the caliper that has taken a knock.
Incoming inspection: When goods-receiving inspection rejects a delivery, your own measurement has to stand up to scrutiny. Having checked the instrument the same day is a strong argument in a discussion with the supplier.
Electronics and precision engineering: Small housings, connectors and sleeves are often measured with the inside jaws. The Ø4 mm ring shows whether the small surfaces are still set correctly.
Why it pays off
The set costs about DKK 2,200. A single faulty batch that has to be scrapped or reworked typically costs far more. In addition, you can limit external calibration to the references themselves and check the rest of the calipers in-house, instead of sending the whole fleet away and going without them for days. That means shorter downtime and fewer uncertain measurements.
Next steps
If you have calipers in daily use without documented checks, the set is an inexpensive way to get started.